QPHY-MOST150o
MOST150 oPHY Serial Data
Operators Manual
Revision B – March, 2012
Relating to the Following Release
Versions:
Software Option Rev. 6.8
MOST150o Script Rev. 1.1
Style Sheet Rev. 1.2
Vue de la page 0
1 2 3 4 5 6 ... 23 24

Résumé du contenu

Page 1 - Versions:

QPHY-MOST150o MOST150 oPHY Serial Data Operator’s Manual Revision B – March, 2012 Relating to the Following Release Versions: •Software

Page 2

10 920456 Rev B connection. Please refer to the QualiPHY Operator’s Manual for explanations on how to use Scope Selector and other QualiPHY funct

Page 3 - 920456 Rev B 3

QPHY-MOST150o Software Option 920456 Rev B 11 4. Click the Configuration button in the QualiPHY main menu: 5. Select a configuration from the p

Page 4

12 920456 Rev B Customizing QualiPHY The predefined configurations in the Configuration screen cannot be modified. However, you can create your

Page 5 - Required Equipment

QPHY-MOST150o Software Option 920456 Rev B 13 Creating Custom Configurations Beginning with any of the pre-loaded configurations, 1. Click on t

Page 6 - USING QUALIPHY MOST150 OPHY

14 920456 Rev B Figure 6 - Variable Setup and Limits Manager windows

Page 7 - 920456 Rev B 7

QPHY-MOST150o Software Option 920456 Rev B 15 QPHY-MOST150o Operation After pressing Start in the QualiPHY menu, the software instructs how to se

Page 8

16 920456 Rev B MOST150 OPHY MEASUREMENT PREPARATION Before beginning any test or data acquisition, the oscilloscope must be warmed for at least 2

Page 9 - Oscilloscope Selection

QPHY-MOST150o Software Option 920456 Rev B 17 QPHY-MOST150O VARIABLES Default Bitrate Specifies whether the bitrate is based on 48 kHz or 44.1 kHz

Page 10 - 10 920456 Rev B

18 920456 Rev B Input Row For Relevant Zi Models Specifies whether to use the A or B inputs when a Zi model oscilloscope with multiple input rows

Page 11 - 920456 Rev B 11

QPHY-MOST150o Software Option 920456 Rev B 19 Eye Diagram Testing 1.2, 2.4, 4.1, 4.3 This test is performed at points SP1, SP2 and SP4 and create

Page 12

2 920456 Rev B LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 578-6020, Fax: (845) 578 5985 Internet: www.

Page 13 - 920456 Rev B 13

20 920456 Rev B Transfer Jitter Testing 1.3, 2.6, 4.2 This test is performed at points SP1, SP2 and SP4. The oscilloscope captures a long signa

Page 14

QPHY-MOST150o Software Option 920456 Rev B 21 Additional tests performed at SP1 Check Bitrate 1.1 The actual bitrate of the signal is measured an

Page 15 - QPHY-MOST150o Operation

22 920456 Rev B B0/B1 Measurements 2.2 This test is provided for informational purposes only. Figure 16 - Oscilloscope Configuration after the B

Page 16 - All Tests [demo]

QPHY-MOST150o Software Option 920456 Rev B 23 Edge Mask Tests 2.3 The specification defines a single mask for the positive signal edges and indivi

Page 17 - QPHY-MOST150O VARIABLES

24 920456 Rev B Transition Time Tests 2.5 Transition times are measured relative to the measured B0/B1 levels and rescaled to give the results in

Page 18 - 18 920456 Rev B

QPHY-MOST150o Software Option 920456 Rev B 3 TABLE OF CONTENTS INTRODUCTION TO QUALIPHY MOST150 OPHY ...

Page 19 - 920456 Rev B 19

4 920456 Rev B FIGURES Figure 1 - Report menu in QualiPHY General Setup...

Page 20 - 20 920456 Rev B

QPHY-MOST150o Software Option 920456 Rev B 5 INTRODUCTION TO QUALIPHY MOST150 OPHY QPHY-MOST150o is an automated test package performing all of t

Page 21 - 920456 Rev B 21

6 920456 Rev B USING QUALIPHY MOST150 OPHY QualiPHY MOST150 oPhy guides the user, step-by-step, through each of the tests described in the specif

Page 22

QPHY-MOST150o Software Option 920456 Rev B 7 QUALIPHY COMPLIANCE TEST PLATFORM QualiPHY is LeCroy’s compliance test framework which leads the us

Page 23 - 920456 Rev B 23

8 920456 Rev B See the QualiPHY Operator’s Manual for more information on how to use the QualiPHY framework. Figure 2 - The Test Report includes

Page 24

QPHY-MOST150o Software Option 920456 Rev B 9 Oscilloscope Option Key Installation The required option keys must be purchased to enable the QPHY-MO