QPHY-PCIE3 Operator’s Manual Revision A – March, 2014 Relating to the following release versions: Software Version Rev. 7.3.0.0 Script PCI_
General Setup The first sub-dialog contains general system settings. These remain in effect for each session, regardless of Standard, until changed
QPHY-PCIE3 Operator’s Manual QualiPHY Test Process Once general system settings are in place, these are the steps for running test sessions. Set Up T
Run Reports The QualiPHY software automates report generation. On the framework dialog, go to General Setup > Report to pre-configure reporting
QPHY-PCIE3 Operator’s Manual Customizing QualiPHY The pre-loaded configurations cannot be modified. However, you can create your own test configurati
Edit Variables The Variable Setup tab contains a list of test variables. See QPHY-PCIE3 Variables for a description of each. To modify a variable:
QPHY-PCIE3 Operator’s Manual Edit Test Limits The Limits tab shows the Limit Set currently associated with the configuration. Any limit set can be as
X-Replay Mode The X-Replay mode window is an advanced (“developer”) view of QualiPHY. The tree in the upper-left frame enables you to navigate to p
QPHY-PCIE3 Operator’s Manual QPHY-PCIE3 Testing Test Preparation Before beginning any test or data acquisition, the oscilloscope should be warmed for
Demo Transmitter Tests This configuration runs all the Transmitter tests (Tests 1.1 – 1.7 below, plus Test 1.4 the Compliance Eye 8 GT/s test) usi
QPHY-PCIE3 Operator’s Manual TEST 1.4 – COMPLIANCE EYE 8 GT/S CEM Compliance Eye 8 GT/s. This test forms an eye diagram using 1,008,000 UI from the
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000, Fax: (845) 578 5985 teledynelecroy.com © 2014 Teledyne LeCroy, Inc.
DISPLAY CONNECTION DIAGRAMS This variable enables/disables the display of DUT connection diagrams during the test process. The default value is Yes
QPHY-PCIE3 Operator’s Manual TX POSITIVE SOURCE This variable allows the user to specify the channel being used to measure Lane A Tx+. TX2 POSITIVE
USE CTLE This variable allows the user to specify whether or not to apply Continuous Time Linear Equalization (CTLE). This variable is set to Yes
QPHY-PCIE3 Operator’s Manual Manual Deskewing Procedures This section is only applicable to the oscilloscope and the cables connecting to the oscillo
• Trigger i. Configure to Source to be FastEdge. ii. Set the Slope to Positive. • Parameter Measurements: i. Set the source for P1 to CX and the
QPHY-PCIE3 Operator’s Manual 10. Save Channel X to M1. • Click File Save Waveform. • Set Save To Memory. • Set the Source to CX. • Set the Dest
Cable Deskewing Without Using the Fast Edge Output The following procedure demonstrates how to manually deskew two oscilloscope channels and cables
QPHY-PCIE3 Operator’s Manual In the previous procedure, we used the default setup of the Skew parameter (which is detecting positive edges on both si
QPHY-PCIE3 Operator’s Manual Table of Contents Introduction ...
Table of Figures Figure 1 - QualiPHY framework dialog and Standard selection menu. ... 5 Figure 2 -
QPHY-PCIE3 Operator’s Manual Introduction About QualiPHY QualiPHY is a highly automated compliance test software meant to help you develop and valida
Required Host Computer System Usually, the oscilloscope is the host computer for the QualiPHY software, and all models that meet the acquisition r
QPHY-PCIE3 Operator’s Manual Installation and Setup QualiPHY is a Windows-based application that can be configured with one or more serial data compl
Set Up Dual Monitor Display Teledyne LeCroy recommends running QualiPHY on an oscilloscope equipped with Dual Monitor Display capability. This allo
QPHY-PCIE3 Operator’s Manual Using QualiPHY This section provides an overview of the QualiPHY user interface and general procedures, using examples