Teledyne-lecroy QPHY-DDR4 Manuel d'utilisateur Page 31

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 73
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 30
QPHY-DDR4 Software Option
Figure 15 - tCK, tCH, tCL and tJIT(duty) Results
In the Measure section:
tCK rise (P1) is the period measurement at Vref (0 mV) of Z1 (differential clock signal) on only the
rising edges. The mean value is the measured value for tCK(avg), rise. The minimum value is
the measured value for tCK(abs), rise, min reported in mtCK(avg). The maximum value is the
measured value for tCK(abs), rise, max reported in mtCK(avg).
tCK fall (P2) is the period measurement at Vref (0 mV) of Z1 (differential clock signal) on only the
falling edges. The mean value is the measured value for tCK(avg), fall. The minimum value is
the measured value for tCK(abs), fall, min reported in mtCK(avg). The maximum value is the
measured value for tCK(abs), fall, max reported in mtCK(avg).
tCH (P3) is the width measure at Vref (0 mV) of Z1 (differential clock signal) on only the high pulses.
The mean value is the measured value for tCH(avg) reported in mtCK(avg). The minimum value
is the measured value for tCH(abs), min reported in mtCK(avg).
tCL (P4) is the width measure at Vref (0 mV) of Z1 (differential clock signal) on only the low pulses.
The mean value is the measured value for tCL(avg) reported in mtCK(avg). The minimum value
is the measured value for tCL(abs), min reported in mtCK(avg).
tJIT(CH) (P5) subtracts the mean of P3 (tCH(avg)) from all of the tCH values. The minimum value is
the measured value for tJIT(CH), min and the maximum value is the measured value for
tJIT(CH), max.
924291 Rev A 31
Vue de la page 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 72 73