QPHY-ENET Operator’s Manual Rev G
Figure 60. Data Amplitude measurement for Splitter calibration
6. Connect AWG to J3 and J4.
7. Set AWG output to 0 V.
8. Remove jumpers from J71, J77, and install jumpers on J17, J19, J91, J92, J99 and J100.
9. On the Fixture Calibration dialog, press Measure (Attenuated Data), wait for it to finish.
10. Press Apply to apply the calibration settings.
Figure 61. Attenuated data measurement for Splitter calibration
Note: The TF-ENET-B fixture attenuation is close to 3.00 dB. This is the default value. You can enter this value on the menu without doing the
calibration.