QPHY-SAS3 Software Option
Figure 19 - Oscilloscope Configuration after the Vpp, VMA, and EQ tests
Shown on this screen:
F2 is Eye Diagram created from the input signal using a 2 pole PLL with a natural frequency of
3.9MHz and a damping factor of .800. This is shown for information only.
F4 is a histogram of the input waveform.
F5 is this histogram converted to a waveform (this is required to perform certain measurements on
the histogram).
F6 is the differential input waveform. This is calculated by subtracting the negative input from the
positive input.
Figure 20 - Vpp, VMA and EQ Results
In the Measure section:
Vpp (P1) is measuring the full range of the F4 histogram. This is the measured result for Tx - Vpp
(5.3.4). This value must be between 850mV and 1.2V in order to pass this test.
LowMode (P4) is the maximum peak from the left side of the F5 waveform. This represents the de-
emphasized low voltage level of the signal.