APPENDIX A: CABLE DESKEWING
There is an automatic cable deskew wizard built in to QPHY-SAS3. When the variable Test Mode is set to
Yes then deskew wizard will prompt the user before the making the first acquisition.
Note: Cables should be deskewed once the temperature of the oscilloscope is stable. The oscilloscope
must be warmed up for about twenty at least a half-hour before proceeding. This procedure should be run
again if the temperature of the oscilloscope changes by more than a few degrees.
Figure 38Startup of Deskew Wizard
If you press Yes the wizard will begin to step through the deskew process. There are two different options
for deskewing: use the MFTP from the DUT or use the Fast Edge from the oscilloscope.
Figure 39Choice of Deskew Method
44 922545 Rev A