Teledyne-lecroy QPHY-ENET Manuel d'utilisateur Page 57

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 78
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 56
QPHY-ENET Software Option
QPHY-ENET Operator’s Manual Rev G
57
1000Base-T Master Jitter - Mode 2 without TX_TCLK
Clause 40.6.1.2.5 of IEEE 802.3-2005 says to measure the jitter between the TX_TCLK and the data at the MDI
(Medium Dependent Interface) that is Jtxout - apply a 5 kHz high-pass filter to the TX_TCLK jitter, measure the
peak-to-peak remaining jitter, and add Jtxout, the result must be less than 0.3 ns. Also, the result of measuring
the peak-to-peak jitter on the TX_TCLK relative to an un-jittered reference must be less than 1.4 ns.
We can assume the Jtxout is small. So, it proposes the following measurements when TX_TCLK is unavailable:
the unfiltered peak-to-peak jitter at the MDI and the peak-to-peak remaining jitter when the unfiltered jitter is
filtered by a 5 kHz high-pass filter.
Since unfiltered jitter measured at the MDI is the sum of TX_TCLK jitter, plus Jtxout, if the result is less than
1.4 ns peak-to-peak then we can be sure that the TX_TCLK jitter is also less than 1.4 ns – in this case we can say
that this measurement “passes.” Keep in mind, this test is more difficult to pass than the test of unfiltered jitter on
TX_TCLK test specified in IEEE 802.3-2005 because the signal on the MDI includes Jtxout. It is not possible to
say with certainty that the DUT should fail this test if the result is more than 1.4ns because it is not possible to
know how much of the measured jitter was Jtxout.
Appendix 40.B-1 (Page 43 of the Gigabit Ethernet Consortium, Clause 40, PMA Version 2,.4:16, June 2006) also
measures the signal at the MDI to approximate the filtered jitter on the TX_TCLK test because the signal on the
MDI is the only signal available.
In the filtered master jitter test in IEEE 802.3-2005 Jtxout is not filtered. The peak-to-peak value of Jtxout
(remember, that is the jitter between TX_TCLK and the signal at the MDI) is added to the filtered TX_TCLK jitter.
In this test, the signal on the MDI (with the TX_TCLK jitter plus Jtxout) is high-pass filtered. Therefore, this test is
easier to pass than the test in 802.3 (because Jtxout is filtered). So, if a device produces a value greater than
0.3 ns, we can say with certainty that it has failed the filtered TX_TCLK test. However, it is not possible to say with
certainty that a DUT passed this test if the filtered jitter at the MDI is less than 0.3 ns. This (like 802.3) is also
because it is not possible to know how much of the measured jitter was due to Jtxout.
So, the larger the contribution from Jtxout, the more this test underestimates the filtered jitter test result, since the
standard test would add the peak-to-peak unfiltered Jtxout to the filtered TX_TCLK peak-to-peak jitter. Appendix
40.B-1 summarizes this in the following
X
Table 8
X
:
Parameter Conformance Limit Result < Limit Result > Limit
Unfiltered TM2 jitter
1.4 ns PASS Inconclusive
Filtered TM2 jitter
0.3 ns Inconclusive FAIL
Table 8. Jitter Test Mode Limits
Perform the test by doing the following:
1. Make the DUT generate the Test Mode 2 signal.
2. Connect the MDI of the DUT to J64 of the TF-ENET text fixture.
3. Connect oscilloscope channel 2 and channel 3 as described in
X
Table 9. Jumper configuration for each
pair measurements
X
(as follows) to test a given pair (A, B, C, D). See
X
Figure 45
X
for an example
connection diagram.
4. On the oscilloscope menu, select Analysis Ethernet tests. Set up the test and press the Set Up and
Start Test button. Both filtered and unfiltered measurements are made on the same data.
Vue de la page 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 77 78

Commentaires sur ces manuels

Pas de commentaire