QPHY-PCIE Testing
Test Preparation
Before beginning any test or data acquisition, warm the oscilloscope for at least 20 minutes.
Calibration is automatic under software control and no manual calibration is required. This
procedure will be run again if the oscilloscope temperature changes by more than a few degrees.
Required Test Modes
The QPHY-PCIE script requires that you place the DUT in the required test modes. The script
will prompt you to do so before each test, but it is recommended that you ensure the DUT is
capable of being placed in the required test modes before beginning testing.
Physical Setup
See the description of each test configuration for specific set up information.
QPHY-PCIE Test Configurations
Test configurations include variable settings, limit sets, and test selections.
For detailed descriptions of calibration and testing, please refer to the Teledyne LeCroy PCIE
Gen1 and Gen2 test procedures on PCISIG.com and the official test specification, PCI Express
Card Electromechanical Specification, Rev. 3.0.
2.5 GTs Clock Jitter Test
This configuration runs only Test 1.6 Clock Jitter for 2.5Gbps signals. Variations exist for
different source configurations: differential inputs on Input A or Input B, and single inputs on
Input A or Input B. This test requires the use of the PCI Express Clock Phase Jitter Test
Software, which can be downloaded from:
http://www.pcisig.com/specifications/pciexpress/compliance/compliance_library
Demo
This configuration runs all the Transmitter tests using waveforms saved on the oscilloscope in
D:\PCIe\Demo and the default limit set, Compliance Limits. This configuration is meant to easily
demonstrate the QPHY-PCIE capabilities when live signals cannot be tested. You will see
dialogs and connection diagrams similar to what you would see during an actual test.
Empty Template
This configuration includes all tests and default variable settings specified in the PCIe standard.
It is meant to serve as a basis for creating custom configurations. The limit set in use is PCIe
2.1. Copy and save it with a unique name, then follow the steps in Customizing QualiPHY
to
modify the new configuration.
Signal Quality Test (CEM), Add-in Card
This configuration is meant to run all of the transmitter tests on a PCIe Add-In card. All the
variables are set to default, except Device Type is set to “Add-In” and CleanClock is set to “Yes”.
The limit set used is PCIe 2.1. Both 2.5Gbps and 5.0Gbps Signal Quality tests are run.